Dynatech MP-7000 Specifications Page 57

  • Download
  • Add to my manuals
  • Print
  • Page
    / 68
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 56
57
Pressure Meters
Ordering information
DPM4-1H Model 1H (± 350 mmHg)
DPM4-1G Model 1G (-700 to 5000
mmHg)
DPM4-2H Model 2H (± 350 mmHg,
Press, Temp, Flow, RH)
DPM4-2G Model 2G (-700 to
5000 mmHg, Press, Temp, Flow, RH)
Included accessories
MANUAL Users Manual
BE-UNVSL-IEC320C14 Battery Eliminator
Power Cord (country specific)
9 Volt Alkaline Battery
Optional accessories
17192 Soft-Sided Carrying Case
14611 PT-100 Temperature Probe
14612 PT-1000 Temperature Probe
14610 Expansion Chamber
14622 Tubing Kit with Inflation Bulb
17449 RS-232 Cable
Temperature Probe PT-100 and PT-1000
PT-100 Operating range -200 °C to 750 °C (-328 °F to 1382 °F)
Accuracy ± 0.13 °C @ 100 °C (0.23 °F at 212 °F)
± 0.1 °C @ 0 °C (0.18 °F @ 32 °F)
± 0.2 °C @ 100 °C (0.36 °F @ 212 °F)
PT-1000 operating range -200 °C to 750 °C (-328 °F to 1382 °F)
Accuracy 0.3 °C (0.5 °F)
Model 2G or 2H Note: It is possible to compensate for the sea level and calibrate for offsets.
Pressure measurement
Operating range 380 mmHg to 825 mmHg
Accuracy ± 2 % of reading
Units of measure mmHg, mBar, cmH
2
O, psi, inHg, inH
2
O, kg/cm
2
, and kPa
Relative humidity Note: An integrated sensor in the instrument determines relative humidity measurements.
Operating range 12 % RH to 95 % RH
Accuracy ± 3.5 % of reading ± 2 % @ 25 °C (77 °F)
Gas flow Note: Gas flow measures with an embedded sensor with 11 calibration points to compensate
non-linearity: calibration constants are stored in firmware.
Gas compatibility Air, N
2
, O
2
, CO, NO, CO
2
, N
2
O, NO
2
Operating range -750 ml/min to 750 ml/min
Accuracy ± 1 % of range or ± 5 % of reading
Gas flow units ml/min or sccm (Standard Cubic Centimeter per Minute)
Peak flow test Peak flow is captured in the unit selected for flow.
A reset key allows to restart the test.
Model 1G, 1H, 2G and 2H
Leak test and peak test
Leak test Leak rate is computed in the unit selected for pressure over 15,
30, 45 or 60 seconds
Peak test Peak pressure is captured in the unit selected for pressure.
A reset key allows to restart the test.
Temperature
Operating 15 °C to 35 °C (59 °F to 95 °F)
Storage 0 °C to 50 °C (32 °F to 122 °F)
General information
Power 9 V alkaline battery RG9 or battery eliminator
Battery life > 7 hours
Dimensions (LxWxH) 156 mm x 94 mm x 34 mm (6.1 in x 3.7 in x 1.3 in)
Weight 0.4 kg (0.9 lb) with battery
DPM4
Parameter Tester
Specications
riešenia na presné meranie
http://www.elso.sk
Page view 56
1 2 ... 52 53 54 55 56 57 58 59 60 61 62 ... 67 68

Comments to this Manuals

No comments