39Multi-Function Simulators
ProSim 4
Vital Signs Simulator
Arrhythmia
Atrial fibrillation Coarse or fine
Premature ventricular contraction Left ventricular
Ventricular tachycardia 160 BPM or 200 BPM
Ventricular fibrillation Coarse or fine
Transvenous pacer pulse 75 BPM, left arterial, 3 mV amplitude on lead II, accuracy ± 10 %, 1.0 ms width
2nd degree AV block Type 1
3rd degree AV block 3rd degree AV block
Asystole Asystole
ECG performance testing
Amplitude 1 mV
Square wave 60 ms at 2 Hz
Respiration
Rate 0 (OFF), 10 BrPM to 100 BrPM in 10 BrPM steps
Impedance variations (Δ Ω) 1 Ω
Accuracy delta ± (10 % + 0.05 ohm)
Baseline 500 Ω to circuit common, giving 1000 Ω between any two leads
Accuracy baseline ± 5 %
Respiration lead LA or LL (default)
Invasive blood pressure
Channels 1 electrically isolated from all other signals
BP output Circular DIN 5-pin
Input/output impedance 300 Ω ± 10 %
Exciter input range 2 to 16 V peak
Exciter-input frequency range DC to 5000 Hz
Transducer sensitivity 5 µV/V/mmHg
Pressure accuracy ± (1 % of setting + 1 mmHg)
Accuracy guaranteed for dc excitation only
Static pressure 0 mmHg, 80 mmHg, 160 mmHg, and 250 mmHg
Dynamic waveforms Synchronization To ECG heartrate
Chambers simulated and systolic/diastolic pressure:
Type IBP (arterial) IBP (left ventrical)
Adult 60/30 60/0
Adult 120/80 120/0
Adult 150/100 150/0
Adult 200/150 200/0
Neonatal 35/15 35/0
Neonatal 70/40 70/0
Non-invasive blood pressure
Pressure units mmHg
Manometer (pressure meter) Range 10 mmHg to 400 mmHg
Resolution 0.1 mmHg (for display purposes)
Accuracy ± (1 % reading + 1 mmHg)
Pressure source Inflation bulb or device under test